UKAS Lab 0625  •  ISO/IEC 17025:2017
Get a Quote

UKAS Lab 0625 | ISO/IEC 17025:2017 | Southampton

UKAS calibration for Southampton

An audit date does not move because a calibration supplier is slow. Labcal calibrates gas and air flow, liquid flow, pressure, temperature, humidity and air velocity instruments at its UKAS-accredited laboratory in Lymington, around 40 minutes from Southampton, with certificates back in 7 working days on average.

7

Working day average turnaround

3-day fast-track available

6

Parameters, one lab

Gas flow, liquid flow, pressure, temp, humidity, velocity

0625

UKAS Lab Number

ISO/IEC 17025:2017 accredited

1

Working day response

We respond to all enquiries within one working day

Southampton logistics

The shortest journey is the one your instrument is not making

Large calibration networks often route instruments through regional hubs. For a Southampton site, a UKAS laboratory covering all six parameters is a short local journey away.

Labcal is based on Ampress Park in Lymington, in the same SO postcode area as Southampton. An instrument sent from the Western Docks, Chilworth or Eastleigh makes a short regional journey rather than a cross-country transfer, so less of your turnaround is spent in transit.

That matters most when you are running without spares. A shorter out-of-service window means fewer duplicate instruments to buy, hire or borrow while the originals are away, and less pressure on the people covering for them.

Map: Southampton to Labcal in LymingtonSchematic map of the Solent showing the road route from Southampton through Totton, Lyndhurst and Brockenhurst to Labcal on Ampress Park, Lymington, around 40 minutes by road.NEW FORESTISLE OF WIGHTTHE SOLENTSOUTHAMPTON WATERM27EastleighTottonChilworthWestern DocksBrockenhurstLyndhurstHytheFawleySouthamptonLABCAL | LAB 0625Ampress Park, Lymingtonc. 40 min
Get directions →
Laboratory

Ampress Park, Lymington SO41 8JU

Turnaround

7 working days on average, 3-day fast-track

Quotes

Like-for-like, within one working day

Collection

For regular or high-volume customers

Southampton’s industries

What Southampton’s industries need calibrated, and why

Southampton’s economy is built around the port, energy and process sites on Southampton Water, research and life sciences, and healthcare. Each one leans on a different mix of measurements.

Port, terminal and marine operations

Wind speed limits govern when cranes and lifting equipment can work, so the anemometers that enforce them need calibration you can defend. Hydraulic and pneumatic pressure gauges on handling equipment sit on the same maintenance schedule.

Energy and process sites on Southampton Water

Process plants depend on pressure gauges, transmitters, flow meters and temperature probes whose readings feed safety and operating decisions. Bringing them under one UKAS schedule gives every inspection the same certificate format.

Research and life sciences

Laboratories around Southampton Science Park and the city’s research base rely on temperature probes, data loggers, humidity sensors and mass flow controllers. Labcal supports customers operating under GMP and ISO 13485.

Healthcare, cleanrooms and building services

Portable anemometers used to commission fume cupboards and cleanrooms are commonly calibrated every year, along with the temperature and humidity instruments that monitor controlled environments.

Testing and calibration laboratories

Hampshire laboratories with their own UKAS scope often need a partner for the parameters they do not hold. Labcal covers all six parameters under one schedule.

Manufacturing and engineering

ISO 9001 sites around Southampton, Totton and Eastleigh need traceable evidence for every instrument that affects product quality. One supplier for all of them means one approved supplier record and one audit trail.

Turnaround at a glance

From your Southampton site to a UKAS certificate

Short transit at both ends means more of the turnaround is spent on calibration, not in a van.

Day 0Dispatch

Courier from your Southampton site, or collection for regular customers.

c. 40 minutesArrival

Instruments are booked in at Ampress Park, Lymington.

7 working daysCalibration

Average from arrival, with a 3-day fast-track by arrangement.

UKAS Lab 0625Certificates

Results and expanded uncertainty (k=2) at every calibrated point.

Time in the laboratory

Standard7 working days (average)Fast-track3 daysDay 0Day 3Day 7

Calibration for Southampton

How calibration works for Southampton businesses

Whether your instruments sit at the port, in a lab on Southampton Science Park or on a production line in Eastleigh or Totton, the process is the same: pack, send, and receive certificates you can hand straight to an auditor.

  1. Tell us what needs calibrating Send instrument types, quantities and the parameters involved through the form below. We respond to all enquiries within one working day with a price and turnaround.
  2. Pack and send from Southampton Use your own courier or ask Labcal to arrange collection. Lymington is roughly 40 minutes by road from Southampton, so next-day services cover the journey comfortably. Collection is available for high-volume or regular customers.
  3. Calibration at UKAS Lab 0625 Instruments are calibrated in our Lymington laboratory against reference standards with a documented traceability chain. Average turnaround is 7 working days from arrival, with a 3-day fast-track by arrangement.
  4. Certificates back to Southampton Every instrument returns with a UKAS certificate showing results and expanded uncertainty (k=2). Keep them for your ISO 9001, ISO 13485, GMP or client audits.

Who in Southampton uses Labcal

Quality and compliance managers who need UKAS evidence for an upcoming audit and one supplier for every parameter.

Procurement teams consolidating calibration spend across several sites in Hampshire and the Solent.

Engineers and metrologists with an instrument that has failed a site check and needs a fast, certified recalibration.

Testing and calibration labs in Hampshire that subcontract parameters outside their own UKAS scope.

Why quality teams choose Labcal

One UKAS schedule covering every parameter you measure

Three reasons quality managers, metrologists and procurement teams in Hampshire and the Solent consolidate their calibration with Labcal.

Six parameters, one accredited supplier

Gas and air flow, liquid flow, pressure, temperature, humidity and air velocity are all held on one UKAS schedule under Lab 0625. Consolidate your programme with a single accredited supplier instead of splitting it across several labs and certificate formats.

Audit-ready, traceable certificates

Every certificate carries documented uncertainty of measurement and a traceability chain back to national standards. Labcal supports customers operating under ISO 9001, GMP and similar quality frameworks that require UKAS-accredited evidence.

Fast, predictable turnaround

Average turnaround is 7 working days from receipt at our Lymington laboratory, with a 3-day fast-track available by arrangement. We can arrange courier collection from Southampton for high-volume or regular customers.

Accredited scope

What Labcal calibrates, and the ranges we are accredited for

Every range below is taken from Labcal’s UKAS schedule of accreditation (Lab 0625, issue 055, August 2026). If your instrument falls inside these ranges, it can be calibrated with a UKAS certificate.

Temperature

UKAS accredited

Digital thermometers, resistance thermometers (PRTs), thermocouples, temperature probes, data loggers, and electrical calibration of temperature indicators and simulators.

MethodAccredited rangeBest expanded uncertainty (k=2)
Resistance thermometers in block or fluid baths-30 °C to +140 °C0.10 °C
Base metal thermocouples in baths-30 °C to +140 °C0.25 °C
Temperature probes in air (environmental chamber)-5 °C to 60 °C0.15 °C
Thermocouple indicator simulation, types T, N, K, J, E-200 °C to +1372 °C (upper limit depends on type)0.50 °C
Thermocouple indicator simulation, types S and R0 °C to 1760 °C0.90 °C
PRT indicator simulation (Pt50 to Pt1000)-200 °C to 850 °C0.16 °C to 0.36 °C
Temperature calibration →

Gas & Air Flow

UKAS accredited

Mass flow meters and controllers, variable area flow meters, gas meters and sonic nozzles. Calibration medium is air; other gases can be used up to 300 l/min. Flow and pressure devices with an electrical output can be calibrated.

MethodAccredited rangeBest expanded uncertainty (k=2)
Prover and reference meter methods0.024 ml/min to 0.5 ml/minQ[1.4 %, 0.0016 ml/min]
Prover and reference meter methods5 ml/min to 10 l/min0.15 %
Prover and reference meter methods0.5 l/min to 2000 l/min0.17 %
Prover and reference meter methods500 l/min to 12,983 l/min0.30 %
Gas and air flow calibration →

Air Velocity

UKAS accredited

Vane anemometers, thermal (hot wire) anemometers and pitot tubes, calibrated in Labcal’s wind tunnels. Anemometers up to 125 mm diameter in the main tunnel; larger instruments in the second facility.

InstrumentAccredited rangeBest expanded uncertainty (k=2)
Thermal anemometers0.1 m/s to 38 m/sQ[0.18 %, 0.030 m/s]
Vane anemometers0.3 m/s to 38 m/sQ[0.22 %, 0.040 m/s]
Pitot tubes2 m/s to 38 m/sQ[0.18 %, 0.030 m/s]
Large anemometers and pitot tubes (second facility)5 m/s to 50 m/sQ[1.7 %, 0.16 m/s] to Q[1.7 %, 0.88 m/s]
Air velocity calibration →

Pressure

UKAS accredited

Pressure gauges, digital pressure indicators, transducers and transmitters, manometers and low differential pressure instruments. Methods consistent with EURAMET CG-17.

MediumAccredited rangeBest expanded uncertainty (k=2)
Gas, gauge (vacuum)-100 kPa to 0 PaQ[0.010 %, 41 Pa]
Gas, gauge (low differential)0 Pa to 3 kPaQ[0.016 %, 0.070 Pa]
Gas, gauge3 kPa to 9 kPaQ[0.016 %, 0.31 Pa]
Gas, gauge9 kPa to 40 kPaQ[0.010 %, 3.4 Pa]
Gas, gauge40 kPa to 500 kPaQ[0.010 %, 41 Pa]
Gas, gauge500 kPa to 8.2 MPaQ[0.010 %, 710 Pa]
Gas, absolute (barometric)80 kPa to 115 kPa0.010 %
Hydraulic, gauge and absolute500 kPa to 140 MPa0.010 % to 0.014 %
Pressure calibration →

Liquid Flow

UKAS accredited

Turbine, Coriolis, electromagnetic and positive displacement flow meters, calibrated gravimetrically with water for mass, volume, flow rate and quantity passed.

MethodAccredited rangeBest expanded uncertainty (k=2)
Gravimetric, water0.5 ml/min to 500 ml/min0.17 %
Gravimetric, water500 ml/min to 2 l/min0.20 %
Gravimetric, water2 l/min to 450 l/min0.14 %
Liquid flow calibration →

Humidity

UKAS accredited

Relative humidity probes, hygrometers, transmitters and combined temperature and humidity instruments, calibrated in an environmental chamber against reference meters.

QuantityAccredited rangeBest expanded uncertainty (k=2)
Relative humidity at 15 °C to 40 °C10 %rh to 90 %rh1.5 %rh
Relative humidity at 10 °C to 15 °C15 %rh to 90 %rh1.5 %rh
Temperature probe built into humidity meter-5 °C to 40 °C0.15 °C
Dew point-15 °C to 40 °C0.18 °C
Humidity calibration →

Uncertainties are the best expanded uncertainties (k=2) on the schedule and depend on the instrument under test. Q[a, b] is the root sum square of a percentage of reading and a fixed value, as written on the UKAS schedule. Where an instrument falls outside an accredited range Labcal can usually offer a traceable (non-UKAS) calibration; say so in your enquiry and we will confirm.

Calibration Services

What can we calibrate for you?

All six parameters are UKAS-accredited under one roof. Choose a parameter to see the full scope, instrument types and how to prepare your equipment for dispatch from Southampton.

Accreditation
UKAS Lab 0625 | ISO/IEC 17025:2017 Independently accredited by UKAS to ISO/IEC 17025:2017.
  • Six parameters on one schedule
  • Expanded uncertainty (k=2) on every certificate
  • Independent of any instrument manufacturer
VIEW UKAS SCHEDULE

Switching labs in Southampton?

Send us your current certificate for a like-for-like quote.

Around half of the quality managers who enquire with us are leaving another calibration provider. We will match your existing scope, parameters and tolerances, then come back with a price and turnaround within one working day. No switching cost, no rework.

Request a Like-for-Like Quote

Common Questions

What Southampton calibration buyers ask most

Straight answers to the questions quality managers and procurement teams ask before choosing a UKAS calibration laboratory.

Is there a UKAS calibration laboratory near Southampton?

Yes. Labcal is a UKAS-accredited calibration laboratory (Lab 0625, ISO/IEC 17025:2017) at Unit 265, Ampress Park, Lymington, Hampshire SO41 8JU, around 40 minutes from Southampton and in the same SO postcode area. Labcal calibrates gas and air flow, liquid flow, pressure, temperature, humidity and air velocity instruments under one schedule of accreditation.

How long does calibration take for a Southampton business?

Average turnaround is 7 working days from the day instruments arrive at the Lymington laboratory. A 3-day fast-track is available by arrangement for instruments needed back urgently, for example ahead of an audit. Because Southampton is close to the lab, courier transit adds very little to the overall time.

Does Labcal calibrate on site in Southampton?

No. All Labcal calibration takes place in the Lymington laboratory, where reference standards, wind tunnels and the environmental chamber are controlled to the conditions on the UKAS schedule. For regular or high-volume Southampton customers, Labcal can arrange collection from site.

How do I get a quote for calibration?

Send your current calibration certificate or an instrument list through the form on this page. Labcal prices the same points and ranges as your existing certificate, so you can compare suppliers on equal terms. We respond to all enquiries within one working day.

What does a Labcal UKAS calibration certificate show?

A Labcal UKAS certificate records the measured results at each calibrated point, the error against the reference and the expanded uncertainty (k=2). It carries the UKAS mark and Lab 0625, identifying the work as calibration within Labcal’s accredited scope.

Does Labcal calibrate instruments from any manufacturer?

Yes. Labcal is an independent laboratory and calibrates instruments regardless of brand. If you are unsure whether a specific instrument falls within the UKAS scope, include the make and model in your enquiry and Labcal will confirm.

What happens if my instrument is outside the UKAS accredited range?

Where an instrument or range falls outside the UKAS schedule, Labcal can usually provide a traceable, non-UKAS calibration instead. The quote states clearly which points are UKAS accredited and which are traceable, so there are no surprises when the certificate arrives.

What is the difference between traceable and UKAS-accredited calibration?

Traceable calibration means the chain of measurement leads back to national standards, but says nothing about the laboratory doing the work. UKAS accreditation adds an independent technical assessment of the laboratory itself, covering its equipment, methods, staff competence and quality management system against ISO/IEC 17025:2017.

Get a Quote

Request a calibration quote

Tell us what you need calibrating. We respond to all enquiries within one working day with a price and turnaround time.

This field is for validation purposes and should be left unchanged.

What do you need? (Required)

Add each type of instrument. Choose "Not sure" at any step and we will work it out with you.

One of our calibration engineers will get back to you by email.

We don't calibrate electrical test equipment, scales and balances, pipettes, gas detectors or dimensional tools.

We respond to all enquiries within one working day. Average turnaround is 7 working days once instruments are received.

Industries we serve in Hampshire and the Solent

UKAS calibration across every regulated sector