UKAS Lab 0625 | ISO/IEC 17025:2017 | Manchester
UKAS calibration for Manchester
An audit date does not move because a calibration supplier is slow. Labcal calibrates gas and air flow, liquid flow, pressure, temperature, humidity and air velocity instruments at its UKAS-accredited laboratory in Lymington, a next-day courier from Manchester, with certificates back in 7 working days on average.
7
Working day average turnaround
3-day fast-track available
6
Parameters, one lab
Gas flow, liquid flow, pressure, temp, humidity, velocity
0625
UKAS Lab Number
ISO/IEC 17025:2017 accredited
1
Working day response
We respond to all enquiries within one working day
Manchester logistics
One box south beats three boxes round the North West
Manchester has calibration labs on its doorstep, and most of them are good at two or three parameters. If your site needs flow, pressure, temperature, humidity and air velocity covered, that usually means three suppliers, three courier runs and three certificate formats.
Labcal holds all six on one UKAS schedule at Ampress Park in Lymington, around 230 miles south. An instrument sent from Trafford Park, Salford Quays, Airport City or Stockport on a next-day service is booked in the following morning. The M6 is a label, not a delay.
One instrument or a hundred, the process is the same: pack it, label it, and it is on the bench in Lymington the next morning. Where it pays off most is the site sending the same kit to three labs to get everything covered. One box south replaces all three, and the audit file gets simpler.
Ampress Park, Lymington SO41 8JU
Next-day courier from Greater Manchester, each way
Like-for-like, within one working day
For regular or high-volume customers
Greater Manchester’s industries
What Greater Manchester’s industries need calibrated, and why
Greater Manchester makes things, tests things and treats people. Advanced manufacturing and aerospace supply chains, one of the UK’s largest life sciences clusters, a hospital estate to match, and a food and drink industry that feeds half the country. Each one leans on a different mix of measurements.
Advanced manufacturing and aerospace supply chain
Trafford Park, Salford and the Bolton and Wigan corridors are full of AS9100 and ISO 9001 sites making parts for someone else’s audit. Pressure gauges, temperature probes and flow meters on test rigs and process lines need certificates a customer’s quality team will accept without a conversation.
Pharma and life sciences
From the Oxford Road corridor and Manchester Science Park out to Alderley Park and Macclesfield, laboratories and manufacturing suites run on temperature probes, data loggers, humidity sensors and mass flow controllers. Labcal supports customers operating under GMP and ISO 13485.
NHS, healthcare and medical devices
Greater Manchester runs one of the largest hospital estates in the country, plus the device makers and sterile services that supply it. Fume cupboards, isolators and pharmacy cleanrooms are commissioned with anemometers, and controlled environments are monitored with temperature and humidity instruments that need an annual certificate.
Food and drink manufacturing
Trafford Park, Wigan and Bolton between them bake, brew and pack for the whole country. BRCGS and retailer audits want traceable certificates for the probe thermometers, data loggers, pressure gauges and flow meters on every line, and they want them on time.
Cleanrooms, data centres and building services
The city’s labs, data halls and hospital wards all depend on controlled air. Vane and thermal anemometers, low differential pressure instruments and humidity probes used by commissioning and facilities teams are calibrated in Labcal’s wind tunnels and environmental chamber.
Testing and calibration laboratories in the North West
The North West has more calibration labs than most regions, and most hold a narrow scope. When a customer asks for air velocity, liquid flow or low differential pressure you do not hold, Labcal takes the subcontract under one UKAS schedule and returns a certificate you can issue against.
Turnaround at a glance
From Manchester to a UKAS certificate, and back
One night in transit each way. Everything in between is calibration.
Next-day courier from your Greater Manchester site, or collection for regular customers.
Instruments are booked in at Ampress Park, Lymington.
Average from arrival, with a 3-day fast-track by arrangement.
Results and expanded uncertainty (k=2) at every point, back to Manchester next day.
Calibration for Manchester
How calibration works for Manchester businesses
Whether your instruments sit on a line in Trafford Park, in a lab off Oxford Road or on a ward in one of the city’s hospitals, the process is the same: pack, send, and receive certificates you can hand straight to an auditor.
- Tell us what needs calibrating Send instrument types, quantities and the parameters involved through the form below. We respond to all enquiries within one working day with a price and turnaround.
- Pack and send from Manchester Use your usual next-day courier or ask Labcal to arrange collection. Instruments dispatched from Greater Manchester in the afternoon are booked in at Lymington the next morning. Collection is available for high-volume or regular customers.
- Calibration at UKAS Lab 0625 Instruments are calibrated in our Lymington laboratory against reference standards with a documented traceability chain. Average turnaround is 7 working days from arrival, with a 3-day fast-track by arrangement.
- Certificates back to Manchester Every instrument returns with a UKAS certificate showing results and expanded uncertainty (k=2). Keep them for your ISO 9001, ISO 13485, GMP or client audits.
Who in Manchester uses Labcal
Quality and compliance managers who need UKAS evidence for an upcoming audit and one supplier for every parameter.
Procurement teams consolidating calibration spend across several sites in Greater Manchester and the North West.
Engineers and metrologists with an instrument that has failed a site check and needs a fast, certified recalibration.
Testing and calibration labs in the North West that subcontract parameters outside their own UKAS scope.
Why quality teams choose Labcal
One UKAS schedule covering every parameter you measure
Three reasons quality managers, metrologists and procurement teams in Greater Manchester and the North West consolidate their calibration with Labcal.
Six parameters, one accredited supplier
Gas and air flow, liquid flow, pressure, temperature, humidity and air velocity are all held on one UKAS schedule under Lab 0625. Consolidate your programme with a single accredited supplier instead of splitting it across several labs and certificate formats.
Audit-ready, traceable certificates
Every certificate carries documented uncertainty of measurement and a traceability chain back to national standards. Labcal supports customers operating under ISO 9001, GMP and similar quality frameworks that require UKAS-accredited evidence.
Fast, predictable turnaround
Average turnaround is 7 working days from receipt at our Lymington laboratory, with a 3-day fast-track available by arrangement. We can arrange courier collection from Greater Manchester for high-volume or regular customers.
Accredited scope
What Labcal calibrates, and the ranges we are accredited for
Every range below is taken from Labcal’s UKAS schedule of accreditation (Lab 0625, issue 055, August 2026). If your instrument falls inside these ranges, it can be calibrated with a UKAS certificate.
Temperature
UKAS accreditedDigital thermometers, resistance thermometers (PRTs), thermocouples, temperature probes, data loggers, and electrical calibration of temperature indicators and simulators.
| Method | Accredited range | Best expanded uncertainty (k=2) |
|---|---|---|
| Resistance thermometers in block or fluid baths | -30 °C to +140 °C | 0.10 °C |
| Base metal thermocouples in baths | -30 °C to +140 °C | 0.25 °C |
| Temperature probes in air (environmental chamber) | -5 °C to 60 °C | 0.15 °C |
| Thermocouple indicator simulation, types T, N, K, J, E | -200 °C to +1372 °C (upper limit depends on type) | 0.50 °C |
| Thermocouple indicator simulation, types S and R | 0 °C to 1760 °C | 0.90 °C |
| PRT indicator simulation (Pt50 to Pt1000) | -200 °C to 850 °C | 0.16 °C to 0.36 °C |
Gas & Air Flow
UKAS accreditedMass flow meters and controllers, variable area flow meters, gas meters and sonic nozzles. Calibration medium is air; other gases can be used up to 300 l/min. Flow and pressure devices with an electrical output can be calibrated.
| Method | Accredited range | Best expanded uncertainty (k=2) |
|---|---|---|
| Prover and reference meter methods | 0.024 ml/min to 0.5 ml/min | Q[1.4 %, 0.0016 ml/min] |
| Prover and reference meter methods | 5 ml/min to 10 l/min | 0.15 % |
| Prover and reference meter methods | 0.5 l/min to 2000 l/min | 0.17 % |
| Prover and reference meter methods | 500 l/min to 12,983 l/min | 0.30 % |
Air Velocity
UKAS accreditedVane anemometers, thermal (hot wire) anemometers and pitot tubes, calibrated in Labcal’s wind tunnels. Anemometers up to 125 mm diameter in the main tunnel; larger instruments in the second facility.
| Instrument | Accredited range | Best expanded uncertainty (k=2) |
|---|---|---|
| Thermal anemometers | 0.1 m/s to 38 m/s | Q[0.18 %, 0.030 m/s] |
| Vane anemometers | 0.3 m/s to 38 m/s | Q[0.22 %, 0.040 m/s] |
| Pitot tubes | 2 m/s to 38 m/s | Q[0.18 %, 0.030 m/s] |
| Large anemometers and pitot tubes (second facility) | 5 m/s to 50 m/s | Q[1.7 %, 0.16 m/s] to Q[1.7 %, 0.88 m/s] |
Pressure
UKAS accreditedPressure gauges, digital pressure indicators, transducers and transmitters, manometers and low differential pressure instruments. Methods consistent with EURAMET CG-17.
| Medium | Accredited range | Best expanded uncertainty (k=2) |
|---|---|---|
| Gas, gauge (vacuum) | -100 kPa to 0 Pa | Q[0.010 %, 41 Pa] |
| Gas, gauge (low differential) | 0 Pa to 3 kPa | Q[0.016 %, 0.070 Pa] |
| Gas, gauge | 3 kPa to 9 kPa | Q[0.016 %, 0.31 Pa] |
| Gas, gauge | 9 kPa to 40 kPa | Q[0.010 %, 3.4 Pa] |
| Gas, gauge | 40 kPa to 500 kPa | Q[0.010 %, 41 Pa] |
| Gas, gauge | 500 kPa to 8.2 MPa | Q[0.010 %, 710 Pa] |
| Gas, absolute (barometric) | 80 kPa to 115 kPa | 0.010 % |
| Hydraulic, gauge and absolute | 500 kPa to 140 MPa | 0.010 % to 0.014 % |
Liquid Flow
UKAS accreditedTurbine, Coriolis, electromagnetic and positive displacement flow meters, calibrated gravimetrically with water for mass, volume, flow rate and quantity passed.
| Method | Accredited range | Best expanded uncertainty (k=2) |
|---|---|---|
| Gravimetric, water | 0.5 ml/min to 500 ml/min | 0.17 % |
| Gravimetric, water | 500 ml/min to 2 l/min | 0.20 % |
| Gravimetric, water | 2 l/min to 450 l/min | 0.14 % |
Humidity
UKAS accreditedRelative humidity probes, hygrometers, transmitters and combined temperature and humidity instruments, calibrated in an environmental chamber against reference meters.
| Quantity | Accredited range | Best expanded uncertainty (k=2) |
|---|---|---|
| Relative humidity at 15 °C to 40 °C | 10 %rh to 90 %rh | 1.5 %rh |
| Relative humidity at 10 °C to 15 °C | 15 %rh to 90 %rh | 1.5 %rh |
| Temperature probe built into humidity meter | -5 °C to 40 °C | 0.15 °C |
| Dew point | -15 °C to 40 °C | 0.18 °C |
Uncertainties are the best expanded uncertainties (k=2) on the schedule and depend on the instrument under test. Q[a, b] is the root sum square of a percentage of reading and a fixed value, as written on the UKAS schedule. Where an instrument falls outside an accredited range Labcal can usually offer a traceable (non-UKAS) calibration; say so in your enquiry and we will confirm.
Calibration Services
What can we calibrate for you?
All six parameters are UKAS-accredited under one roof. Choose a parameter to see the full scope, instrument types and how to prepare your equipment for dispatch from Manchester.

- Six parameters on one schedule
- Expanded uncertainty (k=2) on every certificate
- Independent of any instrument manufacturer
Switching labs in Manchester?
Send us your current certificate for a like-for-like quote.
Around half of the quality managers who enquire with us are leaving another calibration provider. We will match your existing scope, parameters and tolerances, then come back with a price and turnaround within one working day. No switching cost, no rework.
Common Questions
What Manchester calibration buyers ask most
Straight answers to the questions quality managers and procurement teams ask before choosing a UKAS calibration laboratory.
Is there a UKAS calibration laboratory in Manchester?
Labcal’s laboratory is in Lymington, Hampshire, not Manchester. Manchester customers send instruments on a next-day courier and receive them back the same way. Labcal is UKAS-accredited (Lab 0625, ISO/IEC 17025:2017) for gas and air flow, liquid flow, pressure, temperature, humidity and air velocity under one schedule, which is why Greater Manchester businesses use it alongside, or instead of, labs closer to home.
How long does calibration take for a Manchester business?
Average turnaround is 7 working days from the day instruments arrive at the Lymington laboratory, with a 3-day fast-track by arrangement for instruments needed back urgently. Next-day courier adds one working day in each direction from Greater Manchester, so plan on around 9 working days door to door for a standard job.
Why would a Manchester company send instruments to Hampshire?
Because the distance costs one courier label each way and buys six UKAS parameters under one roof. Most calibration labs within reach of Manchester hold two or three. Whether you are sending a single anemometer or a whole site’s instruments, consolidating with one UKAS laboratory means one courier, one certificate format and one approved supplier record.
Does Labcal collect from Manchester?
For regular or high-volume Greater Manchester customers, Labcal can arrange collection from site. Otherwise, pack the instruments and use your usual next-day courier; Labcal confirms receipt when they are booked in at Lymington.
How do I get a quote for calibration?
Send your current calibration certificate or an instrument list through the form on this page. Labcal prices the same points and ranges as your existing certificate, so you can compare suppliers on equal terms. We respond to all enquiries within one working day.
What does a Labcal UKAS calibration certificate show?
A Labcal UKAS certificate records the measured results at each calibrated point, the error against the reference and the expanded uncertainty (k=2). It carries the UKAS mark and Lab 0625, identifying the work as calibration within Labcal’s accredited scope.
Does Labcal calibrate instruments from any manufacturer?
Yes. Labcal is an independent laboratory and calibrates instruments regardless of brand. If you are unsure whether a specific instrument falls within the UKAS scope, include the make and model in your enquiry and Labcal will confirm.
What is the difference between traceable and UKAS-accredited calibration?
Traceable calibration means the chain of measurement leads back to national standards, but says nothing about the laboratory doing the work. UKAS accreditation adds an independent technical assessment of the laboratory itself, covering its equipment, methods, staff competence and quality management system against ISO/IEC 17025:2017.
Request a calibration quote
Tell us what you need calibrating. We respond to all enquiries within one working day with a price and turnaround time.
We respond to all enquiries within one working day. Average turnaround is 7 working days once instruments are received.